Two-parameter leak estimation in non-invasive ventilation


Title

Two-parameter leak estimation in non-invasive ventilation

Publication Type
Conference Paper
Year of Publication
2017
Conference Name
2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Publisher
IEEE
Conference Location
Jeju Island, South Korea
Publication Language
eng
Citation Key
3153
Abstract

In this paper we present a method for the estimation of leaks in non-invasive ventilation. Accurate estimation of leaks is a key component of a ventilator, since it determines the ventilator performance in terms of patient-ventilator synchrony and air volume delivery. In particular, in non-invasive ventilation, the patient flow is significantly different from the flow measured at the ventilator outlet. This is mostly due to the vent orifice along the tube that is used for exhalation, but also to the non-intentional leaks that occur elsewhere in the circuit (e.g., at the mask). Such leaks are traditionally quantified via a model with two parameters, but only one of them is continually updated – the other is fixed. The new algorithm allows for breath-by-breath update of both parameters. This was made possible by leveraging a model describing the patient respiratory mechanics.